Install TRION/TRION3 module in power analyzer DEWE3-PA8
Release time: 22-11-26 08:52 Number of views: forty-six

stay DEWE3-PA8 power analyzer TRION installed in/ TRION3 Board module




 image

one DEWE3-PA8 Mainframe of power analyzer

2 TRION/TRION3 series power module

3 Insert/pull out handle

4 Module Guidance Label

5 Mounting screws


Installation steps of power analyzer DEWE3-PA8 module:


Step 1. Appropriate ESD precautions must be taken to avoid any unit damage.  

Step 2: Disconnect and unplug all connected cables, including those from DEWE3-PA8 and TRION ™/ TRION3 Series module.  

Step 3: Determine a supported TRION ™/ TRION3 Peripheral slot. Some modules require a TRION STAR slot of the.

Step 4: Remove the unused TRION Filter panel for peripheral devices or STAR slots.  

Step 5: Set TRION ™/ TRION3 The module edge of the module is placed at the top and bottom of the chassis in the module guide.  

Step 6: Set TRION ™/ TRION3 Insert the module into the rear of the chassis until resistance appears.

Step 7: Pull up the upper insert/pull out handle to lock the device

Step 8: Use the mounting screws to install the TRION ™/ TRION3 The front panel is secured to the chassis.

Warning: !   The unused TRION slot cannot remain uncovered! Make sure to reinstall the unused TRION Slot to ensure proper cooling of installed modules. If the module overheats due to lack of filter panel, the warranty is invalid!