Description of sensor power supply of DVC power analyzer
Release time: 22-11-30 11:04 Number of views: twenty-eight
Load capacity:
Output power (sum) | 150 W |
Output current+15VLine (single/total) | 1.5 A/5 A |
Output current - 15VLine (single/total) | 1.5 A/5 A |
Output current+9VLine (single/total) | 0.6 A/0.6 A |
Output current (single connector) | 1.5 A |
PIN pin definition:
Compatible transformer/sensor (partial)
PA-IT-60/65 | ±15 V | |
PA-IT-200/205 | ±15 V | |
PA-IT-400/405 | ±15 V | |
PA-IT-700 | ±15 V | |
PA-IT-1000 | ±15 V | |
PA-IN-1000-S | ±15 V | |
PA-IN-2000-S | ±15 V | |
PA-LF-X | ±15 V | |
SE-CUR-CLAMP-150-DC | ±15 V | |
SE-CUR-CLAMP-200-DC | ±15 V | |
SE-CUR-CLAMP-500-DC | ±15 V | |
SE-CUR-CLAMP-1000-DC | ±15 V | |
SE-CUR-LFR | +9 V | |
Quantity limit of PA-IT-1000 and PA-IN-2000-S
As the maximum total output current is 5A, according to the concurrency coefficient of the applied power system, no more than 4x PA-IT - 1000 or PA-IN-2000-S can be provided under the limit conditions. Since the co-occurrence coefficient is less than 1, the following applications can be satisfied (selected) with the nominal current:
>4x1 phase DC system (4x PAIT1000 or PAIN-2000-S sensor)
>4x2 phase DC system (8x PAIT1000 or PAIN-2000-S sensor)
>2x3 phase AC system+1x1 phase DC system (7x PAIT1000 or PAIN-2000-S sensor)
>1x6 phase AC system+1x1 phase DC system (7x PAIT1000 or PAIN-2000-S sensor)